Artifacts caused by transcranial magnetic stimulation coils and EEG electrodes in T(2)*-weighted echo-planar imaging

Magn Reson Imaging. 2000 May;18(4):479-84. doi: 10.1016/s0730-725x(00)00122-3.

Abstract

We investigated the effects of transcranial magnetic stimulation (TMS) coils and electroencephalographic (EEG) electrodes on T(2)*-weighted echo-planar images (EPI) at 2.0 T (gradient-echo EPI, mean TE = 53 ms, 2x2x4 mm(3)). In comparison with anatomic gradient-echo images (3D FLASH, TE = 4 ms, 1x1x1 mm(3)), T(2)*-weighted EPI acquisitions of a water-filled spherical phantom revealed severe signal losses and geometric distortions in the vicinity of TMS coils. Even remote effects were observed for image orientations perpendicular to the coil plane. EEG electrodes and the fixation gel caused milder localized distortions. In humans, complications were avoided by the large distance between the TMS coil and the cortical surface and when using an EPI orientation parallel to the plane of the coil. It is concluded that T(2)*-weighted EPI studies of human brain function may be performed without distortions caused by TMS coils and EEG electrodes.