Large-scale neuroanatomy using LASSO: Loop-based Automated Serial Sectioning Operation

  • Timothy J. Lee ,

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Software, Supervision, Validation, Visualization, Writing – original draft, Writing – review & editing

    timothy.lee@gatech.edu

    Affiliation Georgia Institute of Technology, G. W. Woodruff School of Mechanical Engineering, Atlanta, GA, United States of America

  • Aditi Kumar,

    Roles Data curation, Formal analysis, Software

    Affiliation Georgia Institute of Technology, G. W. Woodruff School of Mechanical Engineering, Atlanta, GA, United States of America

  • Aishwarya H. Balwani,

    Roles Formal analysis, Visualization, Writing – original draft

    Affiliation Georgia Institute of Technology, School of Electrical and Computer Engineering, Atlanta, GA, United States of America

  • Derrick Brittain,

    Roles Conceptualization, Investigation, Methodology, Supervision

    Affiliation Allen Institute for Brain Science, Seattle, WA, United States of America

  • Sam Kinn,

    Roles Data curation, Software

    Affiliation Allen Institute for Brain Science, Seattle, WA, United States of America

  • Craig A. Tovey,

    Roles Conceptualization, Formal analysis, Methodology

    Affiliation Georgia Institute of Technology, H. Milton Stewart School of Industrial & Systems Engineering, Atlanta, GA, United States of America

  • Eva L. Dyer,

    Roles Formal analysis, Visualization, Writing – original draft, Writing – review & editing

    Affiliations Georgia Institute of Technology, School of Electrical and Computer Engineering, Atlanta, GA, United States of America, Georgia Institute of Technology, Coulter Department of Biomedical Engineering, Atlanta, GA, United States of America

  • Nuno M. da Costa,

    Roles Conceptualization, Funding acquisition, Project administration, Resources

    Affiliation Allen Institute for Brain Science, Seattle, WA, United States of America

  • R. Clay Reid,

    Roles Conceptualization, Funding acquisition, Project administration, Resources, Supervision, Writing – review & editing

    Affiliation Allen Institute for Brain Science, Seattle, WA, United States of America

  • Craig R. Forest ,

    Roles Conceptualization, Funding acquisition, Investigation, Project administration, Resources, Supervision, Writing – original draft, Writing – review & editing

    ‡ These authors are joint senior authors on this work.

    Affiliation Georgia Institute of Technology, G. W. Woodruff School of Mechanical Engineering, Atlanta, GA, United States of America

  • Daniel J. Bumbarger

    Roles Conceptualization, Data curation, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Writing – review & editing

    ‡ These authors are joint senior authors on this work.

    Affiliation Allen Institute for Brain Science, Seattle, WA, United States of America

Abstract

Serial section transmission electron microscopy (ssTEM) is the most promising tool for investigating the three-dimensional anatomy of the brain with nanometer resolution. Yet as the field progresses to larger volumes of brain tissue, new methods for high-yield, low-cost, and high-throughput serial sectioning are required. Here, we introduce LASSO (Loop-based Automated Serial Sectioning Operation), in which serial sections are processed in “batches.” Batches are quantized groups of individual sections that, in LASSO, are cut with a diamond knife, picked up from an attached waterboat, and placed onto microfabricated TEM substrates using rapid, accurate, and repeatable robotic tools. Additionally, we introduce mathematical models for ssTEM with respect to yield, throughput, and cost to access ssTEM scalability. To validate the method experimentally, we processed 729 serial sections of human brain tissue (~40 nm x 1 mm x 1 mm). Section yield was 727/729 (99.7%). Sections were placed accurately and repeatably (x-direction: -20 ± 110 μm (1 s.d.), y-direction: 60 ± 150 μm (1 s.d.)) with a mean cycle time of 43 s ± 12 s (1 s.d.). High-magnification (2.5 nm/px) TEM imaging was conducted to measure the image quality. We report no significant distortion, information loss, or substrate-derived artifacts in the TEM images. Quantitatively, the edge spread function across vesicle edges and image contrast were comparable, suggesting that LASSO does not negatively affect image quality. In total, LASSO compares favorably with traditional serial sectioning methods with respect to throughput, yield, and cost for large-scale experiments, and represents a flexible, scalable, and accessible technology platform to enable the next generation of neuroanatomical studies.

Introduction

Serial section transmission electron microscopy (ssTEM) is the most promising tool for investigating the three-dimensional structure of the brain with nanometer-scale resolution [13]. In recent years, ssTEM studies have provided significant insight into the physiology and neuroanatomy of mammalian and non-mammalian nervous systems with resolution and scope previously not possible [46], (respectively, references I, L, M in Fig 1). From published ssTEM literature, we observe a general trend of increasing neural tissue volume studied over time, exemplifying the scientific interest in the field to study larger and larger volumes of neural tissue, as shown in Fig 1. Yet, a significant challenge remains in the scalability of ssTEM. As the volume of brain tissue to be studied grows larger, does ssTEM remain a viable technology in terms of yield, cost, and throughput?

thumbnail
Fig 1. Volume of neural tissue in ssTEM studies versus publication year.

Each data point represents one journal publication that used ssTEM for neuroanatomical studies. We observe the general trend of increasing neural tissue volume studied over time. The largest neuroanatomical ssTEM study (reference M Zheng, Z., et. al., 207) to date remains an order of magnitude below a cubic millimeter.

https://doi.org/10.1371/journal.pone.0206172.g001

Currently, the ssTEM workflow is composed of five primary steps: (1) bulk tissue processing, (2) serial sectioning, (3) post-staining, (4) TEM imaging, and (5) image segmentation. In recent years, significant advances in bulk tissue processing, post-staining, TEM imaging, and image segmentation have shown the potential scalability of ssTEM [5, 713]. In contrast, traditional techniques and methods for serial sectioning for ssTEM have remained unchanged for nearly 60 years [14, 15]. A recent advancement in serial sectioning technology, called the Automatic Tape-collecting Lathe Ultramicrotome (ATLUM), shows promise for the automation of serial sectioning, but demonstration and characterization for neuroanatomical ssTEM studies remains to be shown [16, 17]. Accordingly, advances in methods for serial sectioning will be necessary for the scalability of ssTEM.

In traditional serial sectioning, an ultrathin (< 40 nm) brain slice (or “section”) is cut with a diamond knife into an attached water reservoir, i.e., “waterboat” (see Fig 2A). Individual sections are typically ultrathin in order to have sufficient out-of-plane resolution during EM tomography. Once cut, the section is suspended on and then transported (e.g., by manually whisking the water’s surface with an eyelash) on the water’s surface to a suitable pickup location [15]. Subsequently, the section is carefully picked up by hand (e.g., by using a loop end-effector; see [1820] for previous loop-based serial section pick-up methods) and placed onto a TEM substrate (e.g., copper, plastic-coated slot-grid), paying close attention to align the section with the substrate aperture (see [15] for further details on traditional serial sectioning methodology.) Thus, for traditional serial sectioning, the manual pickup of sections must be repeated for each section without error. From prior work, experienced investigators typically experience 1–3% section loss; with regards to cost, each TEM grid costs ~40 cents; with regards to throughput; each cycle of ultrathin sectioning—transporting, picking up, and placing the section—takes approximately 2 min [4, 5, 13, 15]. While these metrics on a section-by-section basis may seem reasonable, it is the multiplication (or “scaling”) of these values by tens of thousands of serial sections that prohibits large ssTEM neuroanatomical studies.

thumbnail
Fig 2.

Traditional serial sectioning (left) as compared to LASSO (right). (a). Sections are cut on an ultramicrotome using a diamond knife and slide into an adjoining water-filled waterboat, where they float on the water surface. Using a TEM grid held by forceps, a skilled user picks up section(s) from the waterboat onto a TEM substrate, e.g., grid (Fig 2A, inset). Scale bar: 10 mm. (b) For LASSO, sections are fabricated in an identical manner as in Fig 2A, using the same ultramicrotome, diamond knife, and waterboat. From the waterboat, sections are picked up using a loop end-effector, actuated via a robotic system composed of three orthogonal linear axes. Held in the loop end-effector by surface tensions forces (Fig 2B, inset, top right), the section is placed onto microfabricated silicon nitride substrates (Fig 2B, inset, bottom right). Multiple sections are placed onto the same substrate, with each section having its own imaging aperture; a set of substrates compose a “batch,” (e.g., 4 substrates = 1 batch, as shown in Fig 2B, inset, bottom right). Scale bar: 100 mm.

https://doi.org/10.1371/journal.pone.0206172.g002

In the field of industrial engineering, batch processing, i.e., the production of goods in quantized groupings (or “batches”), is a common methodology for yield, cost, and throughput optimization for scaling manufacturing processes [2124]. First introduced by the Toyota Motor Company in the late 1930’s to compete with continuous processing methods (e.g., the assembly line), batch processing enables high-yield, low-cost, high-throughput production [24]. We set out to explore whether batch processing may be an effective methodology for scaling serial sectioning for large neuroanatomical ssTEM studies.

We introduce conceptually, experimentally, and mathematically an alternative method for large-scale serial sectioning, termed Loop-based, Automated Serial Sectioning Operation (LASSO). In our methodology, individual sections are picked up from the waterboat and placed onto TEM substrates using robotic tools for accurate and repeatable, rapid serial sectioning. Batches of sections are placed onto custom microfabricated substrates, reducing overall handling and imaging time of sections. In total, we present a flexible, scalable, and accessible technology platform for serial sectioning to enable the next generation of large-scale neuroanatomical ssTEM studies.

Materials and methods

In LASSO, sections are cut on a conventional ultramicrotome with a diamond knife into an adjoining waterboat, as depicted in Fig 2B. From the waterboat, sections are picked up using a loop end-effector, controlled by three orthogonal linear actuators. Held in place by water surface tension forces within the loop (Fig 2B, inset, top right), the sections are placed onto a microfabricated silicon/silicon nitride TEM substrate (Fig 2B, inset, bottom right). Each substrate holds more than one section with each section having its own imaging aperture. A set of substrates composes a “batch,” (e.g., 4 substrates = 1 batch, as shown in Fig 2B, inset, bottom right). Consecutive sections are never placed on the same batch; therefore, the probability of consecutive section loss is minimized.

Experiment methods

Experimentally, microfabricated substrates were manufactured using conventional semiconductor processing techniques, outlined below in Fig 3. A detailed fabrication protocol is provided in the Supporting Information, see S1 Protocol. A single tissue block was manually trimmed to a cross sectional of ~1 mm x ~1 mm and placed into an ultramicrotome (Leica UC7). Four microfabricated substrates were placed on a hotplate (VWR), adjacent to the ultramicrotome, comprising 160 imaging apertures. The hotplate was set to 95 ºC to enable rapid (< 30 s) drying of sections once placed onto the substrates. The ultramicrotome was set at cutting speed of 0.1 mm/s. Sections were transported away from the knife-edge via a puff of air, delivered manually through an air needle placed adjacent to the waterboat.

thumbnail
Fig 3. Overview of fabrication of silicon/silicon-nitride (Si/SiN) TEM substrates.

Wafers are initially deposited with low-stress silicon nitride (100 nm-thick) followed by photolithography, and plasma etching. In practice we fabricated eight substrates on one 100-mm diameter wafer. See Supplemental Information for fabrication plan details.

https://doi.org/10.1371/journal.pone.0206172.g003

A 3-axis robotic system (Zaber Technologies), positioned adjacent to the ultramicrotome, was used to transport sections between the diamond knife water boat (Diatome) and a microfabricated substrate, using a platinum-iridium wire loop (3 mm diameter) end-effector (TedPella) rigidly affixed to the terminal axis. Section pickup was conducted using manual control of the actuator system via an Xbox controller while placement of the sections onto the microfabricated substrates was automated, following a pre-specified array pattern matching that of the microfabricated substrate apertures. Custom Python-based software was used to interface with the Xbox controller and the actuator system.

After serial sectioning, sections were picked up onto the substrates using the loop end-effector and 3-axis robotic system, with no consecutive sections being placed on the same substrate. A movie of the section pick up and placement process is provided in the Supporting Information, see S1 Movie. Between each section, the ultramicrotome was manually paused to allow for section pickup and placement; once pickup and placement were complete, the ultramicrotome sectioning was resumed. Section pickup and placement was conducted until all 160 apertures were occupied with sections or until a 2-hour imposed time limit to prevent effects due to water evaporation.

Video recording of serial section pickup and placement was conducted to analyze processing throughput and section placement accuracy, repeatability, and yield. From these videos, a MATLAB-script was used to automatically identify and compute each section centroid. Sections that had separated into two or more pieces (but were successfully placed onto the substrate) were excluded from our accuracy and repeatability analysis to maintain a consistent section centroid definition. Yet, these partial sections were picked up and placed successfully. In determining yield, we defined section loss as sections lost due to broken substrate apertures, due to failed section pickup (e.g., loop end-effector damages the section during pickup, or partial sections were unrecoverable), due to failed section transport (e.g., surface tension within loop end-effector breaks during actuator movement), or due to failed placement (i.e., section is not released from the loop end-effector). Thus we define single section loss rate, p, as the probability of damaging or losing one section due to any of these occurrences. We note, in our yield calculation, we include sections that were partially misplaced over the TEM substrate imaging aperture but were successfully placed onto the substrate.

Substrates were imaged using a JEOL EX-II TEM with a camera array system (see [13] for more information regarding the camera array system). An alternative, commercially available loadlock (Voxa) was used to fixture and position the microfabricated substrates during imaging (see [25] for further details regarding the custom loadlock.)

To compare the image quality of the microfabricated films (SiN) used in LASSO with conventional TEM grids (Luxel), we used two standard metrics to quantify the detectability of neuroanatomical structures (e.g., synaptic vesicles). First, the edge spread function (ESF) was computed by manually annotating synaptic vesicles in ITK-Snap on four separate images (2 SiN, 2 Luxel), drawing a line from the vesicle exterior to the interior, and then measuring the change in pixel intensities across the vesicle boundaries [26]. Using derivative-based change-point detection, the change-points in the ESF were obtained and a line was fit to all points between the identified change points. Thus, the slope of this line, i.e., the mean roll-off of the ESF, provides a quantitative measure of the sharpness of edges for neuroanatomical structures of interest in our images [27]. To quantify image contrast, the Michelson contrast was calculated for vesicle interior versus vesicle exterior points. Again, the vesicle interior and exterior pixel intensities were obtained from manually annotated vesicles (ITK-Snap). Then, the mean contrast value across all annotated vesicles was computed, giving a quantitative metric of the detectability of neuroanatomical structures in our images.

Mathematical modeling

Yield modeling

As a statistical model for predicting yield, we let each section pickup and placement event be a binomial random variable, where n is the number of sections to be processed and p be the probability of failure, i.e., damaging or losing one section. To successfully reconstruct a cubic millimeter of neural tissue with 40 nm-thick sections, 25,000 consecutive sections must be cut and imaged with zero consecutive section loss. Sections must be 40 nm thick or less to resolve distal neuronal processes that often are ~100 nm thick, thereby spatially sampling above the Nyquist frequency [3]. From prior literature, we can expect a single-section loss rate lower bound of 1% [4, 5, 13]. The probability of losing two consecutive sections then, assuming P(1 lost section) = p = .01, is

Assuming n = 25,000 and a binomial probability distribution, the probably of losing 1 or more pairs of consecutive sections is The yield, or probability of success assuming failure criteria described above, for an experiment with 25,000 sections traditionally processed is thus 1–0.92 = 8%. Thus, with traditional serial sectioning, large-scale neuroanatomical studies are expected to be impractical. For LASSO, a “batch” of sections will comprise c*m sections, where m is the number of substrates per batch and c is the number of sections per substrate. In total, there are k batches where (1) and n is the total number of section to be processed. Let us assume the loss rate of a single substrate is p’. This is likely an overestimate of the substrate loss rate, since as substrates become larger and easier to handle, we expect fewer substrates to be lost. Within one batch, we will mandate that no two consecutive sections be placed on the same substrate. Therefore, we must lose two or more substrates to lose two consecutive sections. This probability, P, can be calculated as (2) This represents the probability of one failed batch. Furthermore, we can calculate the probability of one or more failed batches out of the total number of batches, k, as (3) Thus, Eqs 13 compose a framework by which to optimize yield by modulating m, the number of substrates per batch, and c, the number of sections per substrate.

Throughput modeling

The total time to collect neuroanatomical datasets can be analyzed on a section-by-section basis. For traditional serial sectioning, the total time for data collection can be written as (4) where timaging is the time to image one section, tpickup, manual is the time to manually pick up and place one section onto a grid, tload time, traditional is the time to load one grid into the TEM, and n is the total number of sections to be processed.

For LASSO, the total time for data collection can be written as (5) where tpickup, robotic is the time to robotically pick up and place one section onto a substrate, tload time, LASSO is the time to load one substrate into the TEM.

Cost modeling

For large sections, each TEM grid typically holds one section. As a result, the total cost for large-scale neuroanatomical datasets can be written as (6) where cgrid is the cost of one grid. For LASSO, sections are placed onto microfabricated substrates, which are manufactured on a wafer-by-wafer basis. (7) where s is the number of sections per wafer, and w is the cost of processing one wafer.

Results and discussion

Modeling results

In developing a mathematical model to predict the likelihood of experiment success (or “yield”) for a large-scale (~1 mm3) neuroanatomical study, we implemented a binomial probability-based model with parameters taken from previously published literature. Using Eqs 13, we plot in Fig 4 the predicted yield for batch processing 25,000 serial sections, i.e., the likelihood of zero consecutive section loss, as a function of c, the number of sections per substrate, and m, the number of substrates per batch. We observe the highest yield is attained by minimizing m while maximizing c (Fig 4, yellow regime). Practically, this corresponds to large EM substrates that can each hold many sections. While this may maximize yield, this optimal solution would be difficult to implement since typical commercial TEMs are designed to hold one grid (~3 mm diameter) with one section on it. Significant modification of a TEM would be required to accept large (>> 3 mm diameter) substrates. Yet, TEM modification is not without precedent, (see references [6] and [13]); thus, the greatest barrier to attaining this theoretical optimum is likely cost. This represents one extreme of our model: a few substrates (m ~ 10), each holding many sections (c ~ 103–104) maximizes the predicted yield. At the opposite extreme, i.e., having many substrates per batch (m ~ 104) with a few sections per substrate (c ~ 1), the model predicts the low yield. Intriguingly, this extreme is analogous to traditional serial sectioning, where one section is placed onto its own grid. Thus, our model points towards the theoretical low-yield associated with performing large-scale ssTEM studies via traditional serial sectioning methods (Fig 4, dark blue regime). The step-like behavior of the plot is due to the need for an integer number of batches. Ultimately, this model captures the spectrum on which ssTEM methodologies lie (with current technologies existing only at the extremes), accurately reflects their respective observed yields, and represents a scaffold for further yield optimization.

thumbnail
Fig 4. Predicted experiment yield for a batch process as a function of m, the number of substrates per batch (1 < m < 100) and c, the number of sections per substrate (0 < c < 80).

The upper bounds for c and m represent practical physical limits while the lower bounds represent minimum physical limits, i.e., for a batch process we mandate at least two substrates per batch with at least one section per substrate.

https://doi.org/10.1371/journal.pone.0206172.g004

Furthermore, in modeling ssTEM experiments, we present a mathematical equation that captures the time required to complete a large-scale ssTEM experiment. As shown in Eq 4, in using traditional serial sectioning methods, the total time for data collection scales linearly with the total number of sections to be processed. In order to decrease the total time for data collection, we can either decrease the section load time, imaging time, or pickup time. From prior work, the imaging time can be decreased by implementing a camera array system (e.g., TEMCA [13]). The section pickup time, using traditional serial sectioning methods is slow (~2 min per section, [15]); therefore, in LASSO, by using robotic pickup we decrease the section pickup time, thereby decreasing the total data collection time. We note that tload time, traditionaltload time, LASSO, but expect tload time, LASSO to decrease with further multiplexing and automated machine vision algorithms. Additionally, by imaging sections in batches, we are able to decrease the section load time by reducing the number required of TEM pump-down cycles for substrate interchange; this is encapsulated in Eq 5 by increasing c and/or m. An optimal solution for both yield and throughput may lie in having exactly two substrates, each holding half of the total number of sections. In this case, significant cost would be required to create custom tooling needed for implementation. Adapting technology developed by Hayworth, K., et. al., [16] to ssTEM may be a viable approach to implement this theoretical optimal solution. In total, we find that traditional serial sectioning methods scale poorly with respect to experiment throughput while LASSO, with batch processing and utilization of robotic tooling, can increase experiment throughput.

Lastly, we present two equations to capture the cost associated specifically with the substrates required for traditional serial sectioning methods versus that of batch processing. For traditional serial sectioning, the total cost of substrates scales linearly with the number of sections (see Eq 6). On the other hand, for LASSO, in which sections are processed in batches, we observe in Eq 7 that the cost per section is inversely related to the number of sections per wafer, assuming a wafer-based microfabrication process. Thus, analogous to that of Moore’s law, we predict the cost per section to decrease as the standard wafer size increases, eventually to below the cost of a single grid [28]. In total, these equations demonstrate the feasibility of implementation of LASSO with respect to cost.

In total, our models for yield, throughput, and cost represent a theoretical scaffold for ssTEM experiment optimization. Furthermore as an experimental implementation of LASSO, using Eqs 13 and setting m and c to be 4 and 40, respectively, we predict a ~91% yield, or probability of success, for an ssTEM study with 25,000 sections—an order of magnitude increase in predicted yield using traditional serial sectioning methods. Accordingly, in our experiments, we implement batch processing with m and c to be 4 and 40 respectively.

Experiment results

We fabricated 26 individual silicon/silicon-nitride substrates. A photograph of four substrates (one batch) is shown in Fig 5A. Each substrate contained forty 1.4 mm x 1.4 mm apertures with pitch 1.9 mm in both the x- and y-direction, as shown in Fig 5B. We processed four wafers and obtained an average yield of 81.25% (6.5/8 substrates) for each wafer. We note, while only 23 of the 26 substrates were used in our experiment, the remaining 3 substrates were of usable quality.

thumbnail
Fig 5.

(a) Photograph of four microfabricated TEM substrates, each with forty 1.4 mm x 1.4 mm apertures (pitch: px = py = 1.9 mm) for TEM imaging. Each aperture contains a 100 nm-thick silicon nitride support films. Scale bar: 10 mm. (b) Photograph of a sub-area of a substrate with sections being placed onto the apertures using a loop end-effector. Scale bar: 1 mm. (c) Representative transmission electron micrograph of section sub-area on the microfabricated TEM substrate. Scale bar: 1 μm. (d) Electron micrograph sub-area depicting labeled vesicle interior (red cross-hair), vesicle exterior (red triangle), and connecting line (red) used to measure edge spread function. Scale bar: 10 nm (e) Mean edge spread function across manually annotated vesicle edges for images from sections on SiN (red) (n = 60) and Luxel (blue) (n = 60). We observe no significant difference in the slope of the ESF, indicating comparable sharpness of edges and image quality. (f) Using LASSO, scatter plot of section centroid positions with x- and y-centroid position distributions; plot extents correspond to imaging aperture size. Centroids located within the outlined box (black dashed line) have their entire area contained within the imaging aperture (587/631 sections, 93%). (g) Histogram of single-section pickup and placement time (solid outline) and single-section pickup time, only (dashed outline).

https://doi.org/10.1371/journal.pone.0206172.g005

We picked up and placed 727 sections across 23 substrates, with a yield of 99.7% (727/729). We find our yield to exceed that of prior work (p < .05), using the Fisher’s exact test, to compare LASSO yield to traditional serial sectioning by Lee, W.A., et. al., (3649/3700 sections) [5]. We note that this comparison is difficult to make due to a variety of uncontrolled parameters (e.g., section size, tissue preparation, sectioning conditions), but in attempting to minimize these differences, we choose to compare to this particular work due to similar section cross-sectional area, section thickness, and the use of mammalian cortical tissue processed for electron microscopy in a similar manner (see [5] for details). The two sections lost were due to substrate aperture failure. To this end, patterning the thin-film with out-of-plane stiffening features (e.g., ribs), may reduce aperture failures. Additionally, alternative aperture materials, such as silicon dioxide or graphene, may further reduce substrate aperture failures. As the field moves towards larger volumes of neural tissue, it is likely that a 1.4 mm x 1.4 mm would be insufficient to encompass the entire section within the imaging aperture. Thus, with the flexibility granted by microfabrication techniques—in particular photolithography—larger apertures can be readily created.

Sparse TEM imaging of sections was conducted to verify image quality. A representative transmission electron micrograph of a section placed on the microfabricated TEM substrate is shown in Fig 5C. Qualitatively, we are able to discern cell membranes and identify individual synaptic vesicles. An image sub-area is shown in Fig 5D, depicting manual annotation of one vesicle using methods described previously. Comparing images obtained on SiN compared to that of Luxel, we observe no visible difference in image quality. Furthermore, the mean slope of the computed edge spread function (ESF) was -2.56 +/-1.07 (n = 60) for Luxel (Fig 5E, blue) and -2.98 +/- 1.83 (n = 60) for SiN (Fig 5E, red), with idealized edges having infinite slope. The contrast for the Luxel film was 0.35 +/- 0.17 (n = 462) and 0.30 +/-0.14 for SiN (n = 636). Our analysis demonstrates that the image quality of the SiN film is comparable to Luxel—a conventionally used TEM grid support film. Thus, LASSO does not adversely affect image quality. Additionally, we observe a slightly larger slope in the SiN ESF, indicating sharper edges in the images. Thus, SiN-based substrates may be a suitable substrate for automated segmentation algorithms [29, 30]. While the Luxel support film does exhibit higher contrast, commonly used contrast enhancing methods (e.g., histogram equalization) could be applied to improve the contrast for SiN substrates.

Applying the criteria previously outlined, we analyzed the section placement accuracy and repeatability for 631/729 sections and report accuracy and repeatability of -20 ± 110 μm (x-axis) and 60 ± 150 μm (y-axis), as shown in Fig 5F. We find that 587/631 (93%) sections lie completely within imaging aperture while the remaining 631–587 = 44 sections are partially occluded from the imaging aperture, i.e., roughly 7% (44/631) sections lie outside of the imaging aperture. While traditional definitions of section loss would consider these 44 sections to be “lost,” we assert that this failure modality is easily remedied via fabrication of substrates with larger imaging apertures. From sections analyzed, given the same accuracy and repeatability, all 631 sections would lie within the 3 mm x 3 mm imaging aperture. We note the 96 sections excluded (due to their breaking into multiple pieces) from our accuracy and repeatability analysis (727–631 = 96 sections) may have fragmented due to a variety of causes, but we do not see this as a fundamental barrier for LASSO due to our ability to successfully pick up and place these sections onto TEM substrates as well as developments in segmentation algorithms to recombine fragmented sections in silico [11]. Further study may investigate potential causes for section fragmentation to prevent partial section all together. Potential causes may include loop geometry, hysteresis of the water with respect to the loop, embedding plastic material properties, embedment protocol, or experiment section dry-down conditions.

We report an average cycle time of 43.5 s ± 11.7 s (Fig 5G, solid outline) and an average section pickup time of 18.9 s ± 9.7 s (Fig 5G, dashed outline). As a first order approximation, we observe a section dry-down time of 24.6 s, calculated as the average cycle time minus the average section pickup time. This is likely an over estimation since this value includes the time required for travel to and from the waterboat. In comparing our average section pickup time that of from prior work (~2 min/section [15]), we find that our methodology has decreased the section pickup time by approximately a factor of five while removing the need for a human user with expert dexterity to pick up serial sections. Additionally, as demonstrated in electrophysiology [31], further automation of section pickup and multiplexing could lower the cycle time even further; thus increasing experiment throughput. Use of machine vision algorithms may accelerate the identification of section centroids while implementation of an automated section transport mechanism may obviate the need for a manually actuated air needle. Automated transport of sections away from the knife-edge prior to section pickup may be accomplished via an automated pneumatic system, standing surface acoustic waves [32] or quadrupolar capillary interactions [33].

LASSO, as previously described, is the composition of several independent technologies that together create flexible, scalable, and accessible platform for large-scale ssTEM. Yet, each technology on its own merits its own discussion. As a part of LASSO, we introduce batch processing, an industrial engineering ideology commonplace in large-scale manufacturing settings. While we demonstrate and characterize a batch processing scheme that utilizes robotic tools and microfabricated substrates, these specifics components are not required to implement batch processing. Moreover, as shown from our modeling results, it is arrangement of serial sections in quantized groupings in a non-consecutive order that enables higher yield. This could be done whether using tape-based substrates [16], silicon wafer substrates [17], or glass slides [34]. By distributing the risk of losing consecutive sections by physically placing them on separate substrates, we maximize the overall experiment yield.

Furthermore, in our implementation of LASSO, we used silicon/silicon-nitride substrates. In their current state, these substrates could be amenable to automated multi-parameter analysis by combining ssTEM with other analysis methods such as electrophysiology [31], array tomography [34, 35], and genetic analysis [36]. Alternatively, these substrates could used with other imaging modalities, such as multi-beam scanning electron microscopy (SEM) [37], scanning transmission electron microscopy (STEM) [38], or x-ray microscopy [39], to provide imaging at multiple resolutions. In this manner, coarse rapid imaging could be combined with high-resolution TEM for morphological alignment. Moreover, further work could explore augmentation of our substrates. A promising direction may be the incorporation of micro-channels within the substrate [40] or as a separate PDMS device [41] to enable in situ staining. Additionally, the substrates could be designed for cell-culture to enable on-chip live-cell imaging/electrophysiology followed by live-cell EM imaging [42, 43]. While each of these are potential future directions, LASSO does not depend on the utilization of microfabricated substrates. In our implementation, the substrates could be readily replaced with tape-based substrates, traditional grids, silicon wafers, or other substrates of choice.

While we demonstrate the ability to manipulate ultrathin sections, LASSO is likely compatible with a variety of section thicknesses. Therefore, semi-thin sections (~100 nm) for light microscopy or EM-based tilt tomography as well as thick sections (~10 μm) for focused ion beam scanning electron microscopy (FIBSEM) could be collected and placed onto an appropriate substrate using LASSO.

While this study was limited to 729 sections, we do not expect fundamental barriers for scaling this technology to 103 to 105 sections, given appropriately apportioned batches. While other technologies may achieve high-yield and high-throughput serial sectioning (see [16, 17]), the accessibility of LASSO, due our use of commercially available cameras, linear actuators, open-source, python-based control software, and standard microfabrication techniques, is a favorable alternative for large-scale serial sectioning.

Conclusion

LASSO represents a flexible, scalable, and accessible technology platform to enable the next generation of large-scale neuroanatomical ssTEM studies. From our modeling, we find that LASSO exceeds the yield, the throughput, and potentially, the cost of traditional serial sectioning methods. Moreover, implementing a batch size of four substrates, with each substrate holding forty sections, we predict an order of magnitude increase in yield. Using this prediction, we microfabricated custom substrates with corresponding size and implemented LASSO to quantify the yield and throughput of the methodology. We find our yield 727/729 (99.7%) exceeds that of prior work (two-sided Fisher test, p = .05); sections were placed accurately and repeatably (x-direction: -20 ± 110 μm (1 s.d.), y-direction: 60 ± 150 μm (1 s.d.)) within the imaging aperture. Sparse TEM imaging of sections showed no significant distortion, high-frequency information loss, or substrate-derived artifacts resulting from serial sectioning via LASSO. Regarding throughput, we find our methodology decreased the section pickup time by a factor of five while removing the need for a human user with expert dexterity to pickup serial sections. (Mean cycle time: 43.5 s ± 11.7 s; mean section pickup time: 18.9 s ± 9.7 s). This technology demonstrates a powerful tool for automating serial sectioning—a significant bottleneck for ssTEM neuroanatomical studies. Thus, we envisage LASSO will enable ssTEM physiological and neuroanatomical studies that investigate neural tissue volumes of size previously not possible, thereby bringing significant insight into the field of neuroscience.

Supporting information

S1 Protocol. Fabrication protocol for TEM substrates.

https://doi.org/10.1371/journal.pone.0206172.s001

(PDF)

S1 Movie. Example of section pickup and placement process.

https://doi.org/10.1371/journal.pone.0206172.s002

(AVI)

Acknowledgments

We acknowledge Marc Takeno, Joann Buchanan, Dr. Adam Bleckert, Dr. Agnes Bodor, and Dr. Stephen Smith, and the Allen Institute for Brain Science for their advice, assistance, and use of equipment in this work. We thank Drs. Tom Reid, Machelle Pardue, and Ross Ethier for their time and use of equipment at Georgia Tech. This work was performed in part at the Georgia Tech Institute for Electronics and Nanotechnology, a member of the National Nanotechnology Coordinated Infrastructure, which is supported by the National Science Foundation (Grant ECCS-1542174). Additionally, this work was supported by the Intelligence Advanced Research Project Activity (IARPA) via Department of Interior/Interior Business Center (Dol/IBC) contract number D16PC00004. The U.S. Government is authorized to reproduce and distribute reprints for Governmental purposes not withstanding any copyright annotation thereon. Disclaimer: The views and conclusions contained herein are those of the authors and should not be interpreted as necessarily representing the official policies or endorsements, either expressed or implied, or IARPA, Dol/IBC, or the U.S. Government. Finally, we wish to thank the Allen Institute for Brain Science founder, Paul G. Allen, for his vision, encouragement, and support.

References

  1. 1. Kornfeld J, Denk W. Progress and remaining challenges in high-throughput volume electron microscopy. Curr. Op. Neurobiol. 2018 June; 50:261–267. pmid:29758457
  2. 2. Gay H, Anderson TF. Serial sections for electron microscopy. Science. 1954 Dec 24; 120(3130):1071–3. pmid:13216224
  3. 3. Briggman KL, Bock DD. Volume electron microscopy for neuronal circuit reconstruction. Curr Opin Neurobiol. 2012 Feb; 22(1):154–161. pmid:22119321
  4. 4. Bumbarger DJ, Riebesell M, Rödelsperger C, Sommer RJ. System-wide Rewiring Underlies Behavioral Differences in Predatory and Bacterial-Feeding Nematodes. Cell. 2013 Jan 17; 152(1–2):109–119. pmid:23332749
  5. 5. Lee WCA, Bonin V, Reed M, Graham BJ, Hood G, Glattfelder K, Reid RC. Anatomy and function of an excitatory network in the visual cortex. Nature. 2016 Apr 21; 532:370–374. pmid:27018655
  6. 6. Zheng Z, Lauritzen JS, Perlman E, Robinson CG, Nichols M, Milkie D, et al. A Complete Electron Microscopy Volume Of The Brain Of Adult Drosophila melanogaster. bioRxiv. 2017 May 22.
  7. 7. Reid RC. From Functional Architecture to Functional Connectomics. Neuron. 2012 Jul 26; 75(2):209–217. pmid:22841307
  8. 8. Hua Y, Laserstein P, Helmstaedter M. Large-volume en-bloc staining for electron microscopy-based connectomics. Nat Commun. 2015 Aug 3; 6:7923. pmid:26235643
  9. 9. Tapia JC, Kasthuri N, Hayworth KJ, Schalek R, Lichtman JW, Smith SJ, Buchanan J. High-contrast en bloc staining of neuronal tissue for field emission scanning electron microscopy. Nat. Protoc. 2012 Jan 12; 7(2):193–206. pmid:22240582
  10. 10. Mikula S, Denk W. High-resolution whole-brain staining for electron microscopic circuit reconstruction. Nat. Meth. 2015 Apr 13; 12: 541–546.
  11. 11. Lee K, Zung J, Li P, Jain V, Seung H. Superhuman Accuracy on the SNEMI3D Connectomics Challenge. arXiv. 2017 May 31.
  12. 12. Januszewski M, Kornfeld J, Li PH, Pope A, Blakely T, Lindsey L, et. al.,. High-precision automated reconstruction of neurons with flood-filling networks. Nat. Meth. 2018 Jul 16; 15:605–610.
  13. 13. Bock DD, Lee WCA, Kerlin AM, Andermann ML, Hood G, Wetzel AW, et al. Network anatomy and in vivo physiology of visual cortical neurons. Nature. 2011 Mar 10; 471:177–182. pmid:21390124
  14. 14. Dowell WCT. Unobstructed mounting of serial section. J Ultrastruct Res. 1959 June; 2(4):388–392.
  15. 15. Harris KM, Perry E, Bourne J, Feinberg M, Ostroff L, Hurlburt J. Uniform Serial Sectioning for Transmission Electron Microscopy. J Neurosci. 2006 Nov 22; 26(47):12101–12103. pmid:17122034
  16. 16. Hayworth KJ, Kasthuri N, Schalek R, Lickthman JW. Automating the Collection of Ultrathin Serial Sections for Large Volume TEM Reconstructions. Microsc. & Microanal. 2006 Jul 31; 12(S02):86–87.
  17. 17. Kasthuri N, Hayworth KJ, Berger DR, Schalek RL, Conchello JA, Knowles-Barley S, et. al. Saturated Reconstruction of a Volume of Neocortex. Cell. 2015 Jul 30; 163(3):648–661.
  18. 18. Gay H, Anderson TF. Serial sections for electron microscopy. Science. 1954 Dec 24; 120(3130):1071–1073. pmid:13216224
  19. 19. Barnes BG and Chambers TC. A simple and rapid method for mounting serial sections for electron microscopy. J. Biophys. Biochem. Cytol. 1961 Mar 1; 9(3):724.
  20. 20. Watanabe S, Jorgensen EM. “Visualizing Proteins in Electron Micrographs at Nanometer Resolution.” Methods in Cell Biology. 2012. 111: 283–306. pmid:22857934
  21. 21. Melouk S, Damodaran P, Chang PY. Minimizing makespan for single machine batch processing with non-identical job sizes using simulated annealing. Int J Production Economics. 2004 Jan 28; 87(2):141–147.
  22. 22. Lee CY, Uzsoy R. Minimizing makespan on single batch processing machine with dynamic job arrivals. Int J Prod Res. 1999; 37(1):219–236.
  23. 23. Mendez CA, Cerda J, Grossman IE, Harjunkoski I, Fahl M. State-of-the-art review of optimization methods for short-term scheduling of batch processes. Comp & Chem Eng. 2006 May 15; 30(6–7):913–946.
  24. 24. Womack J, Jones D, Roos D, Massachusetts Institute of Technology. The Machine that Changed the World: Based on the Massachusetts Institute of Technology 5-million dollar 5-year study on the future of the automobile. 1st ed. New York: Rawson Associates; 1990.
  25. 25. Own CS, Murfitt MF, Own LS, Brittain D, da Costa N, Reid RC, Hildebrand DGC, Graham B, Lee WCA. Reel-to-Reel Electron Microscopy: Latency-Free Continuous Imaging of Large Sample Volumes. Microsc. & Microanal. 2015; 21(Suppl. 3) 157–158.
  26. 26. Trägårdh J, Macrae K, Travis C, Amor R, Norris G, Wilson SH, et. al. A simple but precise method for quantitative measurement of the quality of the laser focus in a scanning optical microscope. Journal of Microscopy. 2015 Jul; 259(1):66–73. pmid:25864964
  27. 27. Yushkevich PA, Piven J, Hazlettt HC, Smith RG, Ho S, Gee JC, Gerig G. User-guided 3D active contour segmentation of anatomical structures: significantly improved efficiency and reliability. Neuroimage. 2006 Jul 1; 31(3):1116–28. pmid:16545965
  28. 28. Thompson SE and Parthasarathy S. Moore’s law: the future of Si microelectronics. Materials Today. 2006 June; 9(6):20–25.
  29. 29. Arganda-Carreras I, Turaga SC, Berger DR, Ciresan D, Giusti A, Gambardella LM, et. al. Crowdsourcing the creation of image segmentation algorithms for connectomics. Front Neuroanat.2015 Nov 5; 9:142. pmid:26594156
  30. 30. Beier T, Pape C, Rahaman N, Prange T, Berg S, Bock DD, et. al. Multicut brings automated neurite segmentation closer to human performance. Nat Meth. 2017 Jan 31; 14:101–102.
  31. 31. Wu Q, Kolb I, Callahan BM, Su Z, Stoy W, Kodandaramaiah SB, et. al. Integration of autopatching with automated pipette and cell detection in vitro. J. Neurophysiol. 2016 Oct 1; 116(4):1564–1578. pmid:27385800
  32. 32. Shi J, Mao X, Ahmed D, Colletti A, Huang TJ. Focusing microparticles in a microfluidic channel with standing surface acoustic waves (SSAW). Lab Chip. 2007 Dec 20; 8(2):221–223. pmid:18231658
  33. 33. Lee TJ, Lewallen CF, Bumbarger DJ, Yunker PJ, Reid RC, and Forest CR. Transport and trapping of nanosheets via hydrodynamic forces and curvature-induced capillary quadrupolar interactions. J Colloid Interface Sci. 2018 Dec 1; 531:352–359. pmid:30041112
  34. 34. Micheva KD, Smith SJ. Array tomography: a new tool for imaging the molecular architecture and ultrastructure of neural circuits. Neuron. 2007 Jul 5; 55(1): 25–36. pmid:17610815
  35. 35. Burel A, Lavault MT, Chevalier C, Gnaegi H, Prigent S, Mucciolo , et. al. A targeted 3D EM and correlative microscopy method using SEM array tomography. Deveopment. 2018 May 25. (forthcoming).
  36. 36. Jiang X, Shen S, Cadwell CR, Berens R, Sinz F, Ecker AS, Patel S, Tolias AS. Principles of connectivity among morphologically defined cell types in adult neocortex. Science. 2015 Nov 27; 350(6264):aac9462. pmid:26612957
  37. 37. Schalek R, Lee D, Kasthuri N, Peleg A, Jones T, Kaynig V, Haehn D, Pfister H, Cox D, and Lichtman J. Imaging a 1 mm3 volume of rat cortex using a multi beam SEM. Microsc Microanal. 2016; 22:582–583.
  38. 38. Kuwajima M, Mendenhall JM, Lindsey LF, and Harris KM. Automated Transmission-Mode Scanning Electron Microscopy (tSEM) for Large Volume Analysis at Nanoscale Resolution. PLoS ONE 8(3): e59573. pmid:23555711
  39. 39. Dyer EL, Roncal WG, Prasad JA, Fernandes HL, Gürsoy D, Andrade VD, et. al. Quantifying Mesoscale Neuroanatomy Using X-Ray Microtomography. eNeuro. 2017 Sep 25; 4(5):0195–17.2017.
  40. 40. de Boer MJ, Tjerkstra RW, Berenschot JW, Jansen HV, Burger GJ, Gardeniers JGE, Elwenspoek M, and van den Berg A. Micromachining of Buried Micro Channels in Silicon. J Microelectromech S. 2000 March; 9(1):94–103.
  41. 41. Walker GM and Beebe DJ. A passive pumping method for microfluidic devices. Lab on a Chip. 2002 Aug; 2(3):131–4. pmid:15100822
  42. 42. de Jonge N, Peckys DB, Kremers GJ, Piston DW. Electron microscopy of whole cells in liquid with nanometer resolution. PNAS. 2009 Feb 17; 106(7):2159–2164. pmid:19164524
  43. 43. Nishiyama H, Suga M, Ogura T, Maruyama Y, Koizumi M, Mio K, Kitamura S, and Sato C. Atmospheric scanning electron microscope observes cells and tissues in open medium through silicon nitride film. Journal of Structural Biology. 2010 March; 169(3):438–449. pmid:20079847