Abstract
The effective connectivity of brain networks can be assessed using functional magnetic resonance imaging (fMRI) to quantify the effects of local electrical microstimulation (EM) on distributed neuronal activity. The delivery of EM to specific brain regions, particularly with layer specificity, requires MRI compatible equipment that provides fine control of a stimulating electrode’s position within the brain while minimizing imaging artifacts. To this end, we developed a microdrive made entirely of MRI compatible materials. The microdrive uses an integrated penetration grid to guide electrodes and relies on a microdrilling technique to eliminate the need for large craniotomies, further reducing implant maintenance and image distortions. The penetration grid additionally serves as a built-in MRI marker, providing a visible fiducial reference for estimating probe trajectories. Following the initial implant procedure, these features allow for multiple electrodes to be inserted, removed, and repositioned with minimal effort, using a screw-type actuator. To validate the design of the microdrive, we conducted an EM-coupled fMRI study with a male macaque monkey. The results verified that the microdrive can be used to deliver EM during MRI procedures with minimal imaging artifacts, even within a 7 Tesla (7T) environment. Future applications of the microdrive include neuronal recordings and targeted drug delivery. We provide computer aided design (CAD) templates and a parts list for modifying and fabricating the microdrive for specific research needs. These designs provide a convenient, cost-effective approach to fabricating MRI compatible microdrives for neuroscience research.
- 3D printing
- effective connectivity
- electrical microstimulation
- electrophysiology
- microdrive
- MRI compatible
Significance Statement
We provide designs for a customizable, magnetic resonance imaging (MRI) compatible microdrive capable of positioning various types of probes (e.g., stimulating electrodes, recording electrodes, drug cannulae, or optogenetic fibers) within the brain. The design integrates a cranial implant, penetration grid for guiding probes, and a microdrive body assembly with actuators. A microdrilling technique, which helps reduce implant maintenance and potential imaging artifacts, is described for introducing probes into the brain. Our open-source designs allow for the customization and fabrication of microdrive components to meet the unique demands of specific research projects and various animal models. Microdrives based on these designs can fulfill a variety of research needs within the neuroscience community related to electrical microstimulation, neuronal recording, and local drug delivery.
Introduction
Hemodynamic responses measured using functional magnetic resonance imaging (fMRI) can be used to study the organization of brain networks (Ogawa and Lee, 1990; Kourtzi et al., 2003; Logothetis, 2003). In particular, fMRI signals evoked by local electrical microstimulation (EM) can reveal the effective connectivity of different brain areas (Tolias et al., 2005; Logothetis et al., 2010; Premereur et al., 2015; Dromme et al., 2016; Duffau, 2020; Premereur and Janssen, 2020). An MRI compatible microdrive capable of positioning various types of probes (e.g., stimulating electrodes, recording electrodes, drug cannulae, or optogenetic fibers) can facilitate a variety of fMRI-based studies. Indeed, incorporating neuromodulation approaches, such as EM and pharmacological methods, into non-human primate neuroimaging studies was recently identified as a five-year goal of “unprecedented value” (Milham et al., 2020).
Currently, relatively few commercially or academically available microdrives can support this goal (Table 1), and several factors limit discovery. First, long travel distances may be required to reach a desired brain region. For example, >30 mm of travel is required to reach some ventral brain regions in macaques. However, only a few available MRI compatible microdrives can travel such long distances. Second, a scanner’s bore size can preclude the use of stereotactic manipulators because of the length of the electrode holders. Third, electrode holders can amplify mechanical vibrations from the scanner, resulting in tissue and/or probe damage. Frameless, skull-mounted microdrives provide effective solutions to problems of bore size and mechanical vibrations but generally have limited travel distance (Wilson et al., 2003; Greenberg and Wilson, 2004). To reduce duplicated efforts associated with groups developing their own microdrives de novo (Moeller et al., 2008; McMahon et al., 2014) and facilitate discovery, we designed a novel MRI compatible microdrive that can be customized to support a wide range of experimental needs.
Here, we present a frameless, skull-mounted MRI compatible microdrive that encapsulates a cranial implant, penetration grid, and actuating mechanism capable of positioning various types of probes over long travel distances. To minimize imaging artifacts, all components are MRI compatible. The penetration grid used in conjunction with a contrast agent is visible in structural scans, providing a fiducial reference for estimating penetration trajectories (Kalwani et al., 2009; Dubowitz and Scadeng, 2011; Glud et al., 2017). The penetration grid further serves as a guide for a microdrilling technique that replaces conventional craniotomies with small holes through which probes are introduced into the brain (Rosenberg et al., 2013; Laurens et al., 2016; Chang et al., 2020). Compared with conventional craniotomies and chambers, this technique reduces potential artifacts caused by air-filled spaces, the need for periodic debridement of granulation tissue, and potential deformation of the underlying neuronal tissue (Wilson et al., 2005; Spitler and Gothard, 2008; Premereur and Janssen, 2020).
We tested the microdrive by using it to introduce a stimulating electrode into the striatum of a macaque monkey and delivered EM during an fMRI session in a 7 Tesla (7T) scanner. Cerebral blood volume (CBV) with monocrystalline iron oxide nanoparticles (MIONs) was used to measure the effects of the delivered EM (Vanduffel et al., 2001; Kim et al., 2013). We provide computer aided design (CAD) files to facilitate modification and fabrication (https://osf.io/tnpmk; RRID:SCR_019247), such as adding/removing actuators and scaling components to accommodate different animal models. We anticipate that this cost-effective, customizable microdrive can support a wide range of neuroscience research requiring fine probe positioning.
Materials and Methods
Microdrive architecture and design
The microdrive consists of two major components: (1) the implant and penetration grid; and (2) the microdrive body assembly (Fig. 1A). The implant is fixed to the cranium and serves as an anchor frame for attaching the penetration grid and microdrive body assembly (Fig. 1B). The penetration grid provides a coordinate system for planning trajectories to reach specific brain areas and guides microdrilling for the insertion of guide tubes which protect the probes (e.g., stimulating electrodes, recording electrodes, drug cannulae, optogenetic fibers, etc.). In the presented design, the microdrive body assembly contains four chambers which align with the four quadrants of the penetration grid. Each chamber houses an individual actuator that independently controls a single probe (Fig. 1C).
CAD files for all parts and assemblies were created using SolidWorks (Dassault Systèmes, SolidWorks Corporation). A Viper si2 SLA System (3D Systems) with 0.0025 mm vertical resolution and 0.075 mm beam diameter was used for 3D printing. All 3D-printed components were made of Accura 60 photopolymer resin (3D Systems), which was selected for its chemical resistance and mechanical durability (tensile strength, 5868 MPa; tensile modulus, 2600–3100 MPa). Where necessary, 3D-printed parts were manually threaded with a standard 2–56 tap (56 threads per inch with a pitch of 453 μm) with the aid of a manual mill to maintain alignment. The actuating rod and guide rods (Fig. 1C) were not 3D-printed and required minor machining (as discussed in the following two sections).
Implant and penetration grid
The implant is a square frame (29 × 29 mm) with four vertical tongues for securing the microdrive body assembly (Fig. 2A) and a threaded hole on each edge for securing guide rods (Fig. 2B, yellow circles). The guide rods are used to align the microdrive body assembly and implant during installation and removal, thereby preventing damage to the guide tubes and/or probes (Fig. 1C; also see Fig. 6D). These rods were made of MRI compatible 3/32” grade 5 titanium alloy and were turned on a lathe. To reduce image distortion, the guide rods can be unscrewed and removed after the microdrive body assembly has been secured to the implant. The outer surface of the implant contains a groove (Fig. 2A, black arrow) which serves as an embedding space when securing the implant to the cranium with dental acrylic (described below).
The main portion of the penetration grid is square in shape with anchor points in each corner that fit into the cutout of the implant (Fig. 2C,D). It is secured to the implant using four Nylon screws (Fig. 2B, blue circles). The grid has four quadrants, each with 61 penetration holes (ϕ0.65 mm; Fig. 2D, orange circles), providing a coordinate system for planning trajectories and guiding microdrilling for the insertion of guide tubes and probes. The location, number, and size of the holes can be modified based on experimental needs.
Microdrive body assembly
The microdrive body assembly consists of two main components: (1) the microdrive body and cap; and (2) the actuating mechanism. In the presented design, the microdrive body contains four chambers (Fig. 3A), each of which can house a single actuating mechanism. The chambers are rectangular (9 × 9 × 45 mm) and align with the four quadrants of the penetration grid (Figs. 2C,D, 3B), providing positioning access for probes over a 353 mm2 area. The bottom corners of each chamber have small stopping notches on which the stopper of the actuating mechanism sits (Fig. 3B, orange triangles). This provides a base of support for the actuating mechanism and prevents it from sliding out the bottom of the chamber. Each face of the microdrive body contains a slot along its length for a guide rod (Fig. 3B, red circles, C, red lines). These rods align the microdrive body assembly with the penetration grid and implant to reduce the risk of damaging guide tubes or probes when installing/removing the microdrive body assembly onto/from the implant (see Fig. 6D). Nylon screws are used to secure the microdrive body to the implant via tapped holes on each face of the body (Fig. 3B, green lines, C, green circles and lines) which align with the tongues on the implant (Fig. 2A,B). The microdrive cap (Figs. 1C, 3D, 4A) holds and aligns the actuating mechanisms within the chambers of the microdrive body (Figs. 3D, blue circles, 4A). The cap also has measurement indicators for tracking the rotation of the actuating rods (Fig. 3D, marks extending radially from each actuating rod hole). The top of the microdrive body contains tapped holes in each corner for securing the cap with Nylon screws (Fig. 3B, purple circles, C, purple lines).
The actuating mechanism, which raises and lowers the probe, consists of three parts: (1) a threaded actuating rod; (2) stopper; and (3) coupler (Fig. 4A). The actuating rod has a scalloped end (Fig. 4B, bottom) that inserts into a latching slot in the center of the stopper (Fig. 4C). The stopper is square and fits snuggly within a chamber of the microdrive body. When an actuating mechanism is assembled and loaded into a chamber, the stopper sits on stopping notches at the bottom of the chamber to prevent the actuating mechanism from sliding out through the bottom of the microdrive body (Fig. 3B, orange triangles; also see Fig. 6C). The coupler fits within the chamber above the stopper and additionally threads onto the actuating rod to control the position of an attached probe (Fig. 4D). Because the coupler fits squarely within the chamber, rotating the actuating rod does not rotate the coupler. Instead, the rotation is converted into vertical translation via a screw and nut mechanism. The stopper provides an anchor point for the actuating rod and coupler that allows the actuating rod to freely rotate because of the latching slot (Fig. 4A). In the presented design, the actuating rod and coupler are threaded for 56 threads per inch, which lowers the coupler and attached probe 453 μm for each full counterclockwise turn of the actuating rod (clockwise rotations raise the coupler and probe). A rotating handle (Fig. 4A) attaches to the beveled end of the actuating rod (Fig. 4B, top) to assist with manual rotations. To prevent unwanted vertical translations of the actuating mechanism within the microdrive body chamber, two brass jam nuts (Fig. 4B, in blue) are used as an adjustable base to support an elastic tube (Fig. 4B, in red), which sits between the nuts and the microdrive cap (Fig. 4A). This structure mimics a compressed spring that creates downward pressure on the actuating mechanism from the microdrive cap, such that the stopper remains pressed against the stopping notches. Thus, the stopper and stopping notches prevent unwanted downward translations, while the brass nuts, elastic tube, and microdrive cap prevent unwanted upward translations. We made the actuating rods from an MRI compatible 2–56 threaded brass rod. The scalloped and beveled ends were manually machined.
Penetration holes in the stopper and coupler (Fig. 4C,D, orange circles) align with corresponding holes in the penetration grid. This alignment defines the penetration trajectory of an attached probe. The number, size, and location of these holes can be modified based on experimental needs. In the presented design, the microdrive can be used to individually control up to four probes simultaneously, one in each chamber. The total travel distance of a probe is determined by the length of the microdrive body and actuating rod (39 mm in this design).
Cranial implant
The implant serves as an anchoring point for the microdrive, so it needs to be secured to the cranium. The implant can be secured directly to the cranium with dental acrylic or to a layer of dental acrylic over the cranium (Figs. 1B, 5A; also see Fig. 6A). An advantage of affixing the implant to an acrylic layer, as opposed to the cranium directly, is that it can be readily repositioned within the bounds of the acrylic layer. The outer surface of the implant contains a groove which serves as an embedding space for the dental acrylic (Fig. 2A, black arrow). Once the acrylic hardens, thus securing the implant, the implant can be protected using a cover that attaches with four Nylon screws (not shown here but included with the CAD files).
Estimating the penetration trajectory using a fiducial MRI marker
Estimating the trajectory and confirming the position of a probe within the brain is crucial to the success of an experiment but can be challenging without a stereotactic frame. Even with a stereotactic device, fine measurements of the distance and angle between the penetration point and target area can be difficult and imprecise. Using the holes in the penetration grid as a visible fiducial MRI marker helps mitigate this problem (Fig. 5). Specifically, the penetration grid can be filled with a contrast agent such as povidone-iodine ointment (Fig. 5A), making it visible in the MRI. In some cases, such as targeting small or deep structures, particularly precise estimates of the probe trajectory may be required. Such estimates can be made using a specialized grid (Fig. 5B). This grid has the same shape and guide hole configuration as the penetration grid but is thicker (10 vs 5 mm) and contains four hollow pillars that are perpendicular to the grid surface. By filling this grid and the pillars with a contrast agent, these features can aid in making more precise estimates of penetration trajectories. As illustrated in Figure 5C–E, the fiducial marker provided by the standard penetration grid (Fig. 5A) can also be used to estimate the probe’s trajectory (Kalwani et al., 2009; Dubowitz and Scadeng, 2011; Glud et al., 2017). Here, the structural MRIs were rotated and aligned to the horizontal plane of the penetration grid using 3DSlicer (Kikinis et al., 2014). In this way, penetration trajectories can be determined in a simple coordinate system defined by the penetration grid. As shown in Results, we tested the microdrive by delivering EM to the striatum during an fMRI session. The penetration hole that we used is marked in Figure 5C (black crosshair), and the trajectory is shown on the rotated coronal and sagittal MRI sections in Figure 5D,E (orange lines).
Preparation of the microdrive assembly and its installation
After identifying the desired penetration hole, a microdrilling technique is used to create a minimum opening through the cranium (here, ϕ0.7 mm; Ideal Micro-Drill, Harvard Apparatus; Fig. 6A). To start, the acrylic layer within the implant is cleaned. The penetration grid is then cleaned with isopropyl alcohol and installed on the implant. A sterilized drill bit with a drill stop collar is then prepared with a bore depth that is slightly longer (e.g., 1 mm) than the thickness of the penetration grid (here, 5 mm). The top of the penetration grid serves as a reference for determining drilling depths and guide tube lengths. After the initial drilling, the drill bit length is sequentially incremented (e.g., in 1 mm steps) until it passes through the cranium. Before engaging the drill, a subsequent drill bit is first used to hand-verify whether the previous bit passed through the cranium. Specifically, the drill bit is lowered through the penetration grid until one of two possibilities occur. If the bit stops before the drill stop collar contacts the penetration grid, then the previous bit did not pass through the cranium and drilling should continue. If the drill bit lowers until the drill stop collar contacts the penetration grid, then the previous drill bit passed through the cranium and drilling should cease. The bore depth that passed through the cranium is used to determine the guide tube length. The guide tube should be slightly longer than the final drill bit plus the distance from the top of the penetration grid to the top of the stopper. In that case, when the guide tube and microdrive body assembly are installed (described below), the guide tube punctures the dura but minimally impinges on the brain.
For MRI studies, guide tubes should be made of a non-metallic material (e.g., silica tubing). However, such materials may not be rigid enough to puncture the dura. This challenge can be overcome by adding an intermediate step in which the dura is punctured using a sterile hypodermic needle. Specifically, the dura can be punctured by preparing and inserting a sterile hypodermic needle that is slightly longer than the final drill bit. After removing the hypodermic needle, a non-metallic guide tube can pass through the opening made in the dura.
Once a guide tube is prepared, it is soaked in a cleaning solution and secured (e.g., using super glue or other high-strength adhesive) to the hole in the stopper that matches the identified hole in the penetration grid (Fig. 6B). The guide tube should be flush with the top of the stopper to ensure the intended guide tube length. The probe is then loaded through the guide tube and secured (again using a high-strength adhesive) to the corresponding hole in the coupler (Fig. 6B). To independently control multiple probes, this process can be repeated with multiple actuators (up to four in the presented design). Each actuating mechanism is then loaded into a chamber of the microdrive body, sliding it down until the stopper rests on the stopping notches within the chamber (Figs. 3B, 6C). The loaded actuators are then secured to the microdrive body using the microdrive cap, which attaches to the top of the microdrive body with four Nylon screws (Figs. 1C, 4A). The microdrive body assembly is then ready to be installed on the implant (Fig. 6C).
Before link installing the microdrive body assembly, four guide rods are screwed into the edges of the implant (Figs. 2A,B, 6D, left). The guide rods are used to align the microdrive body with the implant and penetration grid, ensuring that the guide tubes and probes (which are secured to stoppers and couplers, respectively) maintain alignment with the prepared penetration holes. The aligned microdrive body assembly is then slowly slid down the guide rods toward the implant while monitoring that the tips of the guide tubes enter the appropriate holes in the penetration grid (Fig. 6D, middle). Once the microdrive is fully lowered, it is secured to the implant with four Nylon screws (Fig. 6D, right). After this setup is complete, a probe can be manually lowered/raised by rotating the actuating rod via the rotating handle. To use the microdrive in an MRI study, the guide rods are unscrewed and removed through the top of the microdrive cap before scanning to prevent imaging artifacts.
If using an electrode, a direct connection to a head stage amplifier can be made with electrical wires. For stereotrodes or tetrodes, an additional electrical interface board can be attached to the top of the microdrive body (not shown here but included with the CAD files).
Experimental procedures
All procedures were approved by the IACUC at Sungkyunkwan University (SKKUIACUC2019-03-11-1) and were in accordance with the NIH Guide for the Care and Use of Laboratory Animals. An adult male rhesus monkey (Macaca mulatta; weight, 10 kg; age, 7 years) was implanted with an MRI-compatible round-shaped PEEK headpost (Micro Integration Technology) that was secured using ceramic screws (Thomas Recording) and dental acrylic (Unifast Trad). The microdrive implant was secured to a layer of acrylic over the cranium. A postsurgical structural scan was used to confirm the location of the implant and to estimate the penetration trajectory (Fig. 5).
MRI scanning preparation
Based on the estimated trajectory (Fig. 5), the penetration grid was used as a guide to drill two holes through the cranium. One hole was for a stimulating electrode and the other was for a ground wire (Fig. 6A). Sterile hypodermic needles were then used to puncture corresponding holes in the dura. A stimulating electrode and ground wire were prepared and loaded into fused silica guide tubes (OD: 666 μm, ID: 449 μm; Polymicro) that were secured to separate actuating mechanisms (Fig. 6B,C). The microdrive body assembly was then installed on the implant and secured using Nylon screws (Fig. 6D). During the EM session, no contrast agent was applied to the implant or penetration grid. Finally, the stimulating electrode was lowered into the brain and the ground wire was lowered into the epidural space.
MRI images were acquired under anesthesia (isoflurane, ∼1–1.2%). The electrode trajectory was confirmed with structural and blood oxygenation level-dependent (BOLD) imaging (Fig. 7A,B) before the injection of MIONs (10 mg/kg; total MION = 100 mg; Biopal).
fMRI scanning
Experiments were conducted using a 7T MRI scanner (Terra, Siemens Healthineers) with a 28-channel knee coil (inner diameter, 15.4 cm). Structural images were acquired using a magnetization-prepared 2 rapid acquisition gradient echoes (MP2RAGE) sequence (TR = 4.3 s; TE = 2.12 ms; slice thickness = 0.5 mm iso). BOLD and MION-enhanced CBV data were collected using a gradient-echo echoplanar imaging (GRE-EPI) sequence (TR = 1.5 s; TE = 20 ms; slice thickness = 0.9 mm iso; 52 slices) with whole-brain coverage.
Electrodes and EM plan
Two 231 μm diameter Platinum-Iridium electrodes (exposed tip diameter ∼2–3 μm; MicroProbes) were used for stimulation (impedance = 30 kΩ) and grounding (impedance ≤ 500 Ω). Electrical impulses were generated by a stimulator with two isolators (Master-9 and ISO-Flex stimulus isolator, A.M.P.I.) to apply biphasic current pulses. Each EM block was triggered and synchronized with the scanning procedures by TTL signals from the MRI scanner (MATLAB, MathWorks). Stimulation consisted of trains of biphasic cathode-leading currents with a pulse width of 400 μs and a current of 500 μA, repeated at 50 Hz (Fig. 7C, right). Stimulation trains lasted 200 ms and were repeated every second for 30 s. Periods of 30 s of stimulation and 30 s of no stimulation were interleaved (Fig. 7C, left). Each stimulation session consisted of 13 blocks (i.e., 13 periods of stimulation and 13 periods of no stimulation interleaved).
Image processing and data analysis
Analyses were performed using MATLAB and the Canlab SPM-based fMRI toolboxes (https://github.com/canlab/CanlabCore; Woo et al., 2017). Structural images were calculated using the following equation: (1)where A is the first inversion contrast image, is the complex conjugate of A, B is the second contrast image from the MP2RAGE sequence, and · denotes absolute value (Marques et al., 2010). This method improves the contrast of the brain, but also increases noise outside of the brain as well as metal-induced susceptibility artifacts (Fig. 7A). Structural images were then co-registered to the mean functional images, reoriented to the D99 atlas (Reveley et al., 2017), and segmented into gray and white matter. Functional images were re-oriented using the reorientation matrix obtained from the structural re-orientating process. The images were then motion-corrected, normalized, and smoothed with a Gaussian kernel (1.0 mm full-width at half-maximum). High-pass filtering (cutoff frequency = 0.008 Hz) was used to remove low-frequency signal drifts from the fMRI time series. The CBV signal was then inverted because an increase in blood volume lowers the signal and darkens the image intensity.
Activation maps based on average CBV measurements across stimulation blocks were constructed using a general linear model (GLM) with a design matrix that included a regressor for the EM by convolving the stimulation profile with a boxcar hemodynamic response function. Head movement parameters were accounted for by including linear and quadratic realignment parameters based on current and previous volumes. Statistical maps were then overlaid on the D99 monkey brain atlas to show the areas activated by EM. Voxel-wise t-contrast activations on the spatial maps (false discovery rate, q < 0.05) were used to determine significant activations. The temporal pattern of the CBV response for each stimulation block was constructed by averaging CBV time courses across all voxels within a region of interest (ROI) that had GLM β values >6 (corresponding to the 80th percentile of non-zero β values across all analyzed regions). Voxels belonging to an ROI were determined using the D99 atlas. Activations were reported as percent signal change.
The CAD files and a parts list for the MRI compatible microdrive are available here: https://osf.io/tnpmk/. We encourage others to make modifications based on their specific research needs and hope that our designs facilitate neuroscience research by reducing the time and effort necessary to solve microdrive-related technical issues.
Results
To test the functionality of the microdrive, we used it to deliver EM to the dorsal caudate nucleus (dCN), ventral caudate nucleus (vCN), and nucleus accumbens (NA) of a male macaque monkey in an EM-coupled 7T fMRI study. Structural and BOLD images were periodically taken while lowering the electrode into these areas to verify the electrode’s location (Fig. 7A,B). Importantly, we found that the microdrive generated minimal imaging artifacts in the structural scans, even directly below the implant (Fig. 7A). Even in the BOLD images, which are more vulnerable to artifacts than structural images (Murakami et al., 2016), there was relatively little image distortion from the microdrive (Fig. 7B). The penetration grid was not visible in these images because it was not filled with a contrast agent during the EM session. As expected, the electrode was visible in both the structural and BOLD scans (Fig. 7A,B, respectively). It appears as a white metal-induced susceptibility artifact in the structural image that is prominent because of how the image was calculated from the MP2RAGE scans (Eq. 1). Importantly, these images confirm that the microdrive itself introduced minimal imaging artifacts, even at 7T, making it possible to monitor the position of the probe with a high degree of accuracy.
To assess the effects of EM delivered using the microdrive, we applied EM at three locations in the left hemisphere: the dCN, vCN, and NA. At each location, 13 blocks of EM were delivered. The resulting activation maps are shown in Figure 8. We first stimulated the dCN (penetration depth from the dura = 13.14 mm; Fig. 8A, red arrow) and found that activity significantly increased in the ipsilateral striatum, agranular frontal area F2, ventro-lateral prefrontal area (46v), mediodorsal thalamus (MD), and ventral anterior thalamus (VA; Fig. 8B). A significant increase in activity was also observed in the contralateral agranular frontal area F7. In the same imaging session, we lowered the stimulating electrode into the vCN (penetration depth from the dura = 18.57 mm; Fig. 8C, red arrow). During EM of the vCN, we found that activity significantly increased in the ipsilateral striatum, 46v, MD, VA, medial prefrontal area (10mc), intermediate agranular insula area (lai), and orbital prefrontal area (12o; Fig. 8D). The observed activations associated with dCN and vCN EM were consistent with the known anatomic connectivity of the striatum (McFarland and Haber, 2002; Haber, 2016). Lastly, we lowered the stimulating electrode into the NA (penetration depth from the dura = 22.88 mm; Fig. 8E, red arrow). During EM of the NA, we found that activity significantly increased in the ipsilateral striatum, amygdala, and hippocampal complex (Fig. 8F). These observed activations were consistent with the known anatomic connectivity of the NA (Alexander et al., 1986; Friedman et al., 2002; Choi et al., 2017).
To confirm that the temporal pattern of the CBV responses followed the EM schedule, we calculated the time course of activation in select ROIs during dCN and NA stimulation. For stimulation of the dCN, we calculated the activity within the dorsal striatum, which included the stimulation site (Fig. 8G,H, orange). For stimulation of the NA, we calculated the activity within the relatively distal amygdala (Fig. 8G,H, green). In both cases, the pattern of activity was robust and temporally locked to the EM schedule, with the signal increasing after the onset of stimulation (Fig. 8H, gray bars) and decreasing after stimulation ceased (Fig. 8H, white bars). These results confirm that the MRI compatible microdrive could be used to reliably deliver EM to designated brain areas in a 7T MRI environment with minimal imaging artifacts.
Discussion
For many neuroscience experiments, the microdrive is an essential nexus between neuronal activity and data acquisition. To satisfy the specific demands of a study, it is often necessary to customize the microdrive’s design to ensure accurate and reliable control of probes. Such customization introduces additional development time and manufacturing costs, particularly when it relies on conventional machining. A potentially more effective approach to customizing microdrives is to use 3D printing capabilities with sub-millimeter resolution. Indeed, the utility of 3D printing has recently become apparent across a wide range of research and medical applications (Liaw and Guvendiren, 2017; Jamróz et al., 2018; Nagarajan et al., 2018).
Here, we demonstrated the feasibility of using a 3D-printed, MRI compatible microdrive in an EM-coupled fMRI study. We found that the workflow from design to CAD to fabrication with 3D printing was seamless, and that 3D printing was especially efficient for constructing the microdrive’s small components. This advantage can further expedite customization. For example, the presented design includes four chambers to accommodate four independent actuators, but the supplied CAD files can be easily modified to print a microdrive body that houses fewer or more actuators depending on the number of target areas and the brain size. The length of the chamber can also be easily modified to accommodate shorter or longer travel distances. Importantly, the simplicity of the workflow can support an almost immediate response to new experimental demands, which is not always possible with commercial manufacturers.
Microdrilling reduces imaging artifacts and implant maintenance
The microdrive produced no substantial imaging artifacts because it was made of MRI compatible materials and few metallic parts. In addition, the use of a penetration grid and microdrilling technique eliminated the need for a larger craniotomy and conventional chamber which can create an air-filled space that is problematic for imaging and can potentially result in deformation of the underlying neural tissue. To minimize damage to the dura and neural tissue during microdrilling, a drill stop collar was used to ensure precise control of the bore depth. By hand-verifying the bore depth, it is possible to determine the drill bit length that passes through the cranium within the tolerance of the step size that the drill bits are incremented (e.g., 1 mm). We have used variations of this technique for years (Rosenberg et al., 2013; Laurens et al., 2016; Chang et al., 2020), and have not encountered problems with infection. Meticulous cleaning of the acrylic layer before and after drilling, cleaning and sterilizing components, and covering the implant outside of the experiment (CAD files for a cover are provided but not shown) greatly reduce the risk of infection. However, if infection occurs, the local acrylic layer could be removed and the infection treated topically. The acrylic layer could then be replaced. Thus, the microdrilling technique helps reduce imaging artifacts associated with larger craniotomies as well as implant maintenance.
Performance of the microdrive during EM-coupled MRI scanning
The microdrive showed excellent performance in a 7T MRI environment where it was used to position a stimulating electrode and deliver EM. We placed the electrode in the dCN, vCN, and NA, and delivered EM during fMRI scanning to assess the effective connectivity of these areas across the brain. For each stimulation location, we found that the induced activation was largely restricted to the ipsilateral side of the brain, consistent with some previous findings (Tolias et al., 2005; Matsui et al., 2011). However, it is also possible that our imaging methods limited the ability to detect contralateral activations. Indeed, other studies have shown some contralateral activations induced by EM, consistent with anatomic connections across hemispheres (Ekstrom et al., 2008; Moeller et al., 2008; Premereur et al., 2015; Murris et al., 2020). As expected for the EM of striatal regions, the activated areas were known constituents of the basal ganglia circuit (Draganski et al., 2008; Haber and Knutson, 2010). The results thus confirmed that our MRI compatible and 3D-printed microdrive can be used to examine the connectivity of brain networks using EM-coupled whole-brain imaging.
Future applications and current limitations
Here, we demonstrated the feasibility of using our 3D-printed microdrive in an EM-coupled fMRI study. Because of the versatile design of the actuating mechanism, the microdrive can accommodate various types of electrodes for neuronal recordings (e.g., linear arrays, stereotrodes, or tetrodes), cannulae for inactivation or neuropharmacological studies, or optogenetic fibers for precise neuronal modulation. Thus, the microdrive can support a wide range of future applications. In the current design, a probe is advanced by manually rotating an actuating rod. However, studies often require a period in which the probe must be moved frequently while continuously monitoring neuronal activity (e.g., while approaching a target brain area). In such cases, a motorized actuating module would be more efficient than the current manual mechanism. We therefore have plans to incorporate a motorized actuating module with control software as an extension of the existing design. Lastly, our skull-mounted frameless design is potentially suitable as a chronic microdrive for longer-term applications. The future addition of a protective housing could enable the microdrive body to stay attached to the implant for an extended period of time, potentially with freely moving animals (Wilson et al., 2003; Greenberg and Wilson, 2004; Sun et al., 2006).
Footnotes
The authors declare no competing financial interests.
This work was supported by a Wisconsin National Primate Research Center Pilot Research Project Grant (A.R. and B.K.), the Whitehall Foundation Research Grant 2016-08-18 (to A.R.), and the National Institutes of Health Grant EY029438 (to A.R.). Further support was provided by the Institute for Basic Science in Korea Grant IBS-R015-D1, the National Institutes of Health Grant P51OD011106 to the Wisconsin National Primate Research Center, and the Morgridge Institute.